MEASURING INSTRUMENTS
Meggar: Meggar is a portable insulation tester used
for measuring resistance of any insulation in the order of mega-ohm in which an
ohmmeter and a d.c. generator are incorporated in single unit.
1. Minimum 10000 ohms
resistance can be measured by meggar insulation tester.
2.It can not be used in
live line.
3. Since a meggar is driven
through gearing and a centrifugally controlled slipping clutch then the
generator speed and output voltage remains more or less constant even when the
handle speed is variable, therefore there will be no appreciable variation of
reading in slow speed and average speed rotation of the meggar.
4. In an ammeter or a
voltmeter when the pointer moves the controlling or restoring torque is
produced by two spiral hair springs wound in the opposite direction. One end of
each spring is attached to the spindle of the pointer while other end is
attached to some fixed point. When the instrument is not in use, the two
springs are in their natural position without any tension or compression with
the zero position of the pointer. But in case of a meggar the deflecting torque
and the restraining torque of the moving element is directly proportional to
the supply voltage. As the restraining or controlling torque is proportional to
the supply voltage obtained from the generator the use of control springs is
eliminated. So the pointer floats over the scale and does not point to zero
except when the instrument is in operation.
5. An extra terminal known
as ‘Guard’ terminal in addition to the usual ‘Line’ and ‘Earth’ terminals is
provided in the high voltage meggar to guard the instrument against the effect
of surface leakage surrounding the line terminal within the tester itself or
the connection of line terminal on the testing object. The guard terminal connected
to the negative terminal of the generator. If there is any leakage current, it
will be trapped by the guard ring and returned directly to the negative
terminal of the generator without passing through the current (deflecting) coil
of the instrument and thus eliminating erroneous indication.
6. Maximum 50 ohms resistance
can be measured by an earth testing meggar.
Ohmmeter: An ohmmeter is a
moderately accurate direct reading instrument used for the measurement of the
resistance of a circuit connected across their terminals. It consists of a
moving coil milli-ammeter, a dry cell, a fixed resistance and a variable
resistance, all mounted within the instrument.
Types of Ohmmeters used in
practice:
(i) Measurement of resistance
(ii) Point to point continuity test of circuits.
(iii) End to end continuity test of a long cable.
DC voltage is applied with insulation testing meggar because
with a.c. voltage the capacitance effect will give wrong results.
With DC voltage the contact resistance between the buried
electrode and the soil will be slightly higher than with AC due to the effect
of electrolysis set up by the passage of current causing a film of gas to be
liberated round the electrode and also to set up a back e.m.f. To avoid such
effect to back e.m.f. due to electrolysis effect AC voltage is applied with
meggar earth tester.
Frequency meter: A frequency meter is a
direct reading instrument which indicate the supply frequency of a circuit
directly. It has two terminals. The frequency meter is connected across the
supply line just like a voltmeter.
Classification of
frequency meters: Three main classes of frequency meters -
(i) The
vibrating read type instruments depending for their operation upon the
phenomenon of mechanical resonance.
(ii) The
deflectional electrodynamics type instruments employing electrical resonance.
(ii) The
deflectional moving iron type instruments whose action depends upon the
variation of impedance of an inductive circuit with frequency.
Moving iron type frequency meter is commonly used.
Power factor meter: (i) Dynamo-meter type which
may be either of electro dynamic or of ferro dynamic form and (ii) Moving iron
type. Both types are available either for single phase or for three phase load.
There are two sets of coils, namely
(i) Fixed coils also known as current coils and
(ii) Moving coils also known as pressure coils.
There are (i) two
fixed coils connected in series with the load and carry the current in the
circuit under test and (ii) two
moving coils fixed rigidly at an angle at 900 apart are connected
across the supply lines through a resistance and an inductance respectively.
(a) For balanced load there
are (i) two moving coils fixed with
their planes 1200 apart and connected through resistance across two
different phases of the supply circuit, (ii)
two fixed coils connected in series with the third phase carrying the current
in the line.
(b) For unbalanced load there
are (i) three fixed coils being
displaced 1200 apart from each other and connected in series in the
three lines (ii) three star
connected moving coils being displaced 1200 apart from each other
and connected through non inductive resistances to the three phases.
(i) In single phase power
factor meter there are 4(four) terminals, (ii) in three phase three wire power factor
meter there are 8(eight) termi9nals for balanced load and 9(nine) terminals
for unbalanced load. (iii) In three
phase four wire power factor meter there are 9(nine) or 10(ten) terminals.
Potentiometer: Potentiometer is an instrument used for accurate measurement of a
low range unknown e.m.f. or potential difference up to 1.5 volt by balancing it
against a known potential difference produced by a known current passing
through a resistance.
Clip - on meter: It is also known as tong
tester is a portable hand operated C.T. based ammeter which has a laminated
core arranged in such a way that the core can be opened out at a hinged section
by pressing a trigger-like projection to encircle the current carrying
conductor. During operation the current carrying conductor acts as a single
turn primary whereas the secondary is connected across a standard ammeter
conveniently mounted on the handle itself. It is used to measure the line
current flowing through a core of a cable without disconnecting the cable
leads. There is no electrical connection between cable line and measuring
instrument.
Phase sequence indicator: Phase sequence indicator
is nothing but a small poly phase motor having a stator consisting of three
coils and a metal disc rotor having an arrow marked on it which determines the
order of phase rotation in a three phase system. The rotation in the direction
of arrow indicates the correct phase sequence i.e. R -> y -> B. The phase
sequence can be corrected by interchanging any o the two leads.
Animated flasher: When a flasher is used for
sequential operation to give appearance of movement in luminous sign by using
quick succession of gradually varying drawings then it is said to be animated
flasher.
High voltage animation
flasher:
For more complicated control if it is required for spelling and animation
effects in one sub-circuit flasher is usually connected on the HT side since
this enables parts of a sign fed from one transformer to be operated at
different times. Such flasher connected on the HT side is called as the high
voltage animation flasher.
Low voltage animation
flasher:
Where the whole sign is required to flash on and off at predetermined intervals
the flasher unit is usually connected on the LT mains side. Several small
transformers up to a total capacity of 1000 volt-amps may be operated by one
switch. When a big animated luminous sign requires power of more than 1000 volt-amps
then it is divided into several sections each having a separate transformer of
capacity not more than 250 volt-amps and lights up fully in each sequential
operation. In that case flashers area introduced in the LT mains of
transformers of different sections. Such flashers connected on the LT side are
called as low voltage animation flashers.
Photo electric control: It can be used to make
the sign automatically switched on when the level of illumination falls below a
predetermined value and switched off again when the illumination is restored.
No comments:
Post a Comment